Transcreener® Instrument Validation Program
BellBrook Labs has designed an instrument validation program in an effort to align the high standards we have set for our assay to your specific instrument. It is our goal that you be as successful as possible using Transcreener assays in as little time as possible. Our scientists personally test each line of instruments and are able to offer you both confidence that an instrument will perform well with our far red FP assays and a guideline for setting up your instrument.
Look for these stickers on instruments that have successfully met the validation criteria.


Validation Criteria
- 384-Well Format
- Z’-Factor ≥ 0.7 at 10% conversion of 10 μM ATP
- Δ mP ≥ 95 mP at 10% conversion of 10 μM ATP
- Z' and Δ mP specifications to be met using Transcreener ADP2 Assay reagents
- Read time to achieve Z' and Δ mP specifications ≤ 5 minutes
| Transcreener® Assay | |||||||
|---|---|---|---|---|---|---|---|
| ADP2 FP and GDP FP |
ADP2 FI and GDP FI | ADP2 TR-FRET |
AMP/GMP FP | UDP FP | ADP FP (old product) |
||
TECAN |
Safire²™ | √ | √ | √ | data avail | data avail | √ |
| Infinite® M200 | √ | ||||||
| Infinite® F200 | √ | ||||||
| Infinite® F500 | √ | √ | |||||
| Infinite® M1000 | √ | data avail | √ | ||||
PerkinElmer |
EnVision® | √ | √ | √ | √ | √ | |
BMG LABTECH |
PHERAstar | √ | √ | ||||
| PHERAstar Plus | √ | √ | √ | data avail | √ | ||
| FLUOstar Omega | √ | ||||||
| POLARstar Omega | X (96 well) | √ | |||||
Molecular Devices |
Analyst® GT | √ | √ | ||||
| SpectraMax M2 | n/a | data avail | n/a | n/a | n/a | n/a | |
BioTek |
Synergy™ 2 | √ | √ | ||||
| Synergy™ 4 | √ | √ | |||||
| Synergy™ H4 | √ | √ | √ | ||||
= validated on platform
= validated on 96 well systems
= data available upon request
= no data yet
= not applicable